Feature:Electrolyte analyzer 1Auto-mark,absorb,measure,flush,display and print. 2Auto-turn to hold on,economize reagent and using cost. 3Well linearity,high veracity,wide testing range,and long service. 4Simple liquid channel to avoid malfunctions like blockage. 5QC analysis process,which could carry through data processing automatically,state and print the mean value,standard difference and CV. 6Single calibrate modulus for each item,automatic or manual for optional,guarantee test veracity. 7Unique press design for pump structure,more effective for extended pump`s life than traditional strain model. 8Perfect function,easy for operation and maintenance. 9Test the surplus reagent in the whole-course automatically and note to add reagent. 10Save tests automatically after power off. 11Low teagent consumption.Automatic washing sample needle to avoid malfunctions and intercross infections. Technical index:
1Display: 240×128 graphic LCD. 2Test items: KNaCliCaTCapH. 3Measure element: Ion selectivity electrode technology. 4Sample: whole blood,serum,plasm,urine,and so forth. 5Sample Dosage: 60ul300ul. 6Rate: standard 80 samples/hour. 7Printer: Internal graphic thermal printer with character of low noise,high speed and long service life. 8Interface: RS-232. 9Memory: 1000 sample results,refresh itself after fill in. 10Testing range and precision Item | Range | Distinction rate | Precision | K | 0.5015.00 mmol/L | 0.01 mmol/L | CV≤1.0 | Na | 30.0200.0 mmol/L | 0.1 mmol/L | CV≤1.0 | Cl | 20.0200.0 mmol/L | 0.1 mmol/L | CV≤1.0 | Ca | 0.105.00 mmol/L | 0.01 mmol/L | CV≤1.0 | pH | 4.009.00 | 0.01 | CV≤0.5 |
|