    MAIN CHARACTERISTICS

    1. Single bed and single bulb.

    2. Design according to the electric mechanism of middle-size machine. The function is full and reasonable.

    3. The high-tension system is of march-past structure. Equipped with afullwave rectification high-voltage generator and independent X-ray tube unit, with large output power and reliable and stable operation.

    4. Supply voltage V,roentgenoscopy,photograpy KV and steeples adjustable.

    5. Equipped with the manostat for filament of X-ray tube and the space charge compensator as to ensure the stable and accurate output.

    6. Equipped with the capacity limitation and protection circuit and KvmA  sinterlock limitation and protection.

    7.Adopt the digital circuit timer,grade according to R10 prioritycoefficient and can control time exactly

    8. Examining bed is of electirc back rotation and is equipped with vibrating filter-ray device.

    9. The spot film device of luminescent screen can hold the film jacket of various sizes and can solenoid valve locking.

    10. The upright post is of top and bottom track.It moves lightly.

    TECHNICAL PARAMETER

    ITEM

    CONTENT

    INDEX


    Roentgenoscopy


    voltage

    50~95KV

    50~95KV Be adjustable continuously

    current

    0.5~5mA

     

    Photography


    voltage

    50~100KV 50~125KV Be adjustable continuously

     

    current

    small focus 50mA and big focus 50 100 150 200mA five grades together

    spot film

    150mA(fixed)

    time

    0.04~6.3s,23 grades together

     

    High-Voltage generator

    capacity

     20KVA(instantaneous)


    Max. Dc output voltage

    100KV 125KV


    Max. Dc output current

    200mA

    X-Ray tube

     

    model

    XD51-20.40/100 XD51-20.40/125

     

    focus

    1X1mm 2X2mm

    Examining

     

    rotary angle

    +90 o ~ 0 o ~ -10 o

     

    specifi-cation of spot film(stomach and intestines photograpy)

     

    lengthwise 500mm

    transverse ±80mm 

    pressure down 300mm

    TECHNICAL PARAMETER

     

    Examining bed

    specifi-cation of spot film

    127X177.8mm(5X7)

    203.2X254mm(8X10)

    254X304.8mm(10X12)

    279.4X355.6mm(11X14)

    filter-ray device

     

    grid characteristic

    N28

     

    Convergence distance

    fo 100cm

     

    Grid ratio

    r8

    distance between bed face and film

    75mm

    overrunning governor

     

    manual single blade

    Upright post

     adjusting scope (distance between top and bottom track)

    2800-3400mm

    move lengthwise

    (along top and bottom track)

    3100mm

    move transverse

    (along horizontal arm)

     800mm(from bulb focus to the centre of upright post)

    move up and down

    (along upright post)

     930~1830mm(from bulb focus to the centre of upright post)


    Supply power


    voltage

    380V/220V

    frequency

    50Hz ± 0.5Hz

     

    inner resistance

    380V:0.9 Ω ,220V:0.3 Ω