MAGNET SYSTEM

Magnet "C" Open, Permanent ND-Fe-B magnet
ANTI-EDDY Current technology
Field strength 0.4 T
Homogeneity 300 mm× 300 mm× 360 mm ≤ 5ppm
Patient Gap 410mm

GRADIENT SYSTEM
Gradient strength 25 m T/m
Rise time <0.4ms
Slew rate 60m T/m/ms
Cooling method: Air-cooled

SPECTROMETER SYSTEM
Frequency Max 20 MHz
RF Power 6kW
Transmit Coil Flat coil
Receive coil: Quadrature coil, phase array coil optional

IMAGING SYSTEM
Opermating system WINDOWS2000/NT/XP
Imaging software MR Software Package
CPU ≥ 2GHz
Hard Disk ≥ 250G
External storage CD-RW
Monitor Professional monitor ≥ 19
Interface DICOM 3.0 Standard

SEQUENCE
Spin-Echo (SE)
Inversion Recovery (IR)
Fast Spin-Echo (FSE) T1, T2, Proton density weighted
Flow Attenuation Inversion Recovery (FLAIR)
Short Inversion Time Recovery (STIR)
Magnetization Transfer Contrast (MTC)
Magnetic Resonance Cholangiopancreatography (MRCP)
Magnetic Resonance Urography(MRU)
Magnetic Resonance Angiography (MRA)
2D/3D Gradient-Echo (GE)
Diffusion-weighted Imaging (DWI)