제품 설명: |
1 When adding up voltage, total current is less than 2A and total power is 500W;
2 The current and voltage of X-ray tubes are fixed in three groups, they are:
X exposure voltag (KVP) 60 65 70
X-ray current (mA) 0.5 0.85 1.10
3 When adding up voltage, total current is less than 2A and total power is 500W
4 Drain current and earth resistance are in accordant with the stipulates in GB9706.1
5 The three groups of voltage (60KV, 65KV, 70KV) are adjustable.
6 Wireless Remote control in any angles within 10 meters part form the control box
7 The parameter of X-ray generator (bulb tube head)
The parameter of X-ray tubes
A Focus index: 0.3*0.3mm2
B The voltage of the-highest-voltage generator: 90KV
C The highest inverse peak voltage: 90KV
D Filament characteristic: Vf=5V+0.5V, f=3A
E Target surface obliquity: 12
F the cooling manner of bulb tube head: Oil immersion
The parameter of X-ray generator (bulb tube head):
A. Focus-skin distance: 200mm
B. The thickness of filter-ray film: More than 1.5mmAl
C. The leakage rate of X-ray: No more than 0.002mGy/h
(No. 29.204.3 in GB9706.12 prescript that the leakage rate of X-ray is no more than 0.25mGy/h)
D. The manner of limiting beam: Using planigraphic device
Model NO.:
Wall-Mounted Type
Company:
Sinor Dental Equipment Co., Ltd. |