Descripción del producto: |
*12.1" color TFT display, unique contrast adjustment function
*Real man voice alarm, windows style ONE level menu, easy to use
*Rich screen layout selections: Standard, Big font, Trend table, NIBP trend, ECG full lead, oxyCRG
*Powerful storage: Maximum 720 hour graphic and tabular trends of all parameters, up to 10000 NIBP review, up to 12 hours of all waveforms Built-in recorder and battery
*Full lead ECG monitor with arrhythmia analysis, ST calculation and Pace analysis. ECG circuit can make correct measurement against defibrillation, ESU and artifact.
*Double-protect for NIBP overpressure, anti-temperature drift design, operation correctly under severe environment to protect the patient
*Apply new SpO2 technique to gain precise measurement results under low-perfusion and finger shiver condition
*Standard network interface, can connect to Central monitor system
*System software can be upgraded through network
*All-round monitor of adult, pediatric and neonatal.
Parameters: 5-lead ECG, RESP, SpO2, NIBP, TEMP, PR, Network;
Optional parameter: 2-IBP, Recorder, Arrythmia, EtCO2, VGA signal output.
ECG
5-lead or 3-lead selectable
Input: 5-lead: RA; LA; RL; LL; V or R; L; N; F; C
Lead selection: I; II; III; AvR; AvL; AvF; Vx; CAL
Gain selection: 12.5mm/s, 25mm/s, 50mm/s
Accuracy: Bpm or %, whichever is greater
Protection: Withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation
Calibration signal: 1mV %
Operation modes: Diagnostic, Monitor, Surgery
Alarm range: Adult: 15-300bpm
Pediatric/Neonatal: 15-350bpm
Alarm range: -2.0mV~2.0mV
S-T detection: YES, Arrhythmia analysis: YES
Alarm: YES, audible and visual alarm, alarm events recallable
NIBP
Method: Oscillometry
Operation modes: Manual / Automatic / STAT
Measurement unit: MmHg/KPa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement range: Adult: 10-270mmHg
Over-pressure protection: YES
Resolution: 1mmHg
Accuracy: Mean error and standard deviation per ANSI / AAMI SP-10
Alarm: Systolic, Diastolic, Mean
SpO2
Measurement range: 0-100%
Accuracy: Digit (70~100%);
0~69% unspecified
Alarm range: 0-100%
Respiration
Method: RA-LL impedance
Pediatric/Neonatal: 6-150rpm
Alarm range: Adult: 6-120rpm
Pediatric/Neonatal: 6-150rpm
Apnea alarm: YES
Temperature
Resolution: 0.1
Channel: Dual-channel
Alarm range: 0-50
Optional parameter: 2-IBP, Recorder, Arrythmia, VGA signal output.
IBP
Channel: 2 channels
Pressure transducer: Sensitivity: 5( mV/V/mmHg)
Transducer sites: ART, PA, CVP, RAP, LAP, ICP
Alarm range: -10~300mmHg
Net Weight: 5.3kg Volume Weight: 10kg |