Erklären: |
Main Technical Parameter
Infusion pump mechanism Peristaltic mechanism, suitable for all standard infusion sets
Maximum infusion rate 2000 ml/h
Programmable flow rate scope 1-2000 ml/h
Fast infusion rate (Bolus rate) 1000 ml/h
Selectable modes 1. Rate mode 2. Time mode 3. Body weight mode
Rate mode Rate: 1-2000 ml/h
Time mode Time: 1-2000minutes Solution limit: 0.1-999.9 ml
Body weight mode Weight: 0.1-300.0 kg
Drug limit: 0.1-999.9 mg
Solution limit: 0.1-999.9 ml
Dose: 0.1-9999.9
(The maximum limit is limited by weight, drug limit, solution limit, dose and unit. )
Unit: Mg/kg/h, ug/kg/min
KVO rate Selectable from 1 ml/h to 5.0 ml/h (When occlusion in running status, it enter into KVO rate automatically. )
Programmable volume limit 0.1-9999.9 ml
Programmable time scope 0-99 hours and 59 minutes
Accumulated volume 0.1-9999.9 ml
Accuracy ± 5%
Power supply AC power, 220V50Hz, 25 VA;
Battery Rechargeable lithium polymer battery, 7.4V, 1600mAh. Easy to replacement.
Maximum consumption 25VA, can use for more thank 5 hours at the rate of 10ml/h after fully recharging.
Battery recharge When connect the pump to AC power, the battery will recharge automatically. (14 hours to fully recharge ).
Fuse 2* 220V 2A\2*12V 2A
Display and indicator information Infusion rate, preset volume, accumulated volume, battery capacity, AC power supply indicator, completion, occlusion, air bubble, bed No., pressure value and so on.
Status indicator Stop, Running, rinse and KVO.
Alarm function Completion, occlusion, air bubble, low battery, bnormity1 and abnormity 2.
Maximum size of outer shell 120× 140× 190 mm
Maximum weight < 2.5 kg
Classification Type II
Outer shell material ABS plastic |