उत्पाद विवरण: |
Technical parameters
Maximum size of outer shell: 120x140x190mm
Maximum weight: <2.0 kg
Power: (city electricity) 50/60Hz AC85-265V
Fuse: 220v0.2A, 12V2.5A
Maximum power consumption: 25VA, running more than two hours after being charged
Applicable infusion pipe: All ordinary home-made infusion pipe of national standards
Feasible infusion speed range: Infusion speed 1-600ml/h
Infusion increment: 1ml/h
Feasible preset infusion volume limit: 1~9999ml or unlimited
Fast infusion volume: 1.0ml (Fast infusion: 600ml/h)
Accumulated infusion volume display: 1~9999ml
Infusion precision: +/-5%
Intravenous infusion speed: 1ml/h KVO rate
Occlusion pressure value: 4-14 psi
Alarm function:
1)infusion completion,
2) occlusion,
3)low battery,
4)installation error,
5)infusion pipe air bubble,
6)liquid exhaustion
Screen display:
1)infusion speed,
2) preset infusion quantity,
3) accumulated infusion quantity,
4) battery capacity,
5) battery charging indication,
6)AC power indication,
7)infusion completion,
8)occlusion alarm,
9)infusion pipe air bubble,
10)bed number
Battery: Rechargeable lithium battery, 7.4V, 1000mAh
Electromagnetic wave (Hertzian wave) protection: Accord with the stipulation of IEC60601-1-2 (1993 edition)
Storage condition: Environment temperature 5~50 C degree pressure 70~106kPa, relative humidity 30%~90%
Electrical safety: Accord with the stipulation of GB9706.1-1995
Company: