Description du produit: |
1. Monitoring parameters including ECG, RESP, NIBP, SPO2, PULSE RATE and DUAL-TEMP
2. Large screen 12.1" color TFT display
3. Display pixels: 1024x768
4. Multi-waveforms rolling(or renovating) display with adjustable display speed
5. S-T segment synchronized detection and analysis
6. Maximum 96-hour graphic and tabular trends of all parameters
7. 300-set NIBP, HEART RATE, TEMP, SPO2, RESP data storage
8. 20-min ECG real time waveform and parameters storage and review
9. Able to be connected with the central monitor station
10. Inner high-definition thermal dot matrix recorder optional, which can output waveforms and characters
11. Built-in battery
12. ECG:
Input: Whole-lead ECG cable, standard AAMI cable connector
Sweep speed: 12.5, 25, 50mm/s
Bandwidth: 0.05~100Hz(for diagnosis)
0.5~40Hz(for monitor or operation)
Calibration signal: 1mV
Protection: Against electrosurgical interference and defibrillation
Heart Rate:
Measuring range: 30~250bpm
Accurracy: +/-1 bpm
Resolution: 1bpm
Alarm mode: Audible and visual alarm
RESP:
Method: Resistance on chest
Meas. Range: 0~99 bpm
Gain selection: X 1/2, x1, x2, x4
Oxygen Saturation(SPO2)
Meas. Range: 35~100%
Accuracy: +/-2% (70~100%)
35~69% unspecified
NIBP
Method: Oscillometric
Mode: Manual, auto, STAT
Meas. Interval in AUTO mode: 1 ~ 250(min)
Meas. Period in STAT mode: 5 min
Meas. Type: SYS, DIA, MEAN
Meas. Range: SYS: 60 ~ 270mmHg
DIA: 30 ~ 230mmHg
MEAN: 40~250mmHg
Resolution: 1 mmHg
Over-pressure protection: YES (Adult 300mmHg; Child 220mmHg)
Temperature
Meas. Range: 0~50C
Resolution: 0.1 C
Accuracy: +/- 0.2 C
Operation Environment
Temperature: 0~45C
Humidity: 30~85%(non-condensing)
Power option
AC power: 100~240V, 50~60Hz, 10VA
Company:
Nanaxxon Holdings Ltd. (Jiangsu) |