Atomic Absorption Spectrophotometer (WFX-210)
- Product Name:
Atomic Absorption Spectrophotometer (WFX-210)
- Category: Others
- Package: Carton or boarding ark
- Price: Get Latest Price
- Product Detials:
Atomic Absorption Spectrophotometer (WFX-210)
1. Innovated flame analysis technique
2. The patented flame analysis technique adopting oxygen-rich air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950ºC, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost. Oxygen-rich flame will not pollute the environment and is not harmful to human bodies. It is a break-through in flame AAS analysis.
3. Popular Czerny-Turner grating monochromator with 1800l/mm provides higher resolution.
4. All plastic pre-mix spray chambers with high anti-corrosion ability.
5. Accurate and reliable graphite furnace analysis
6. Light-controlled graphite furnace power supplier, adopting FUZZY-PID temperature control technique and dual curve working mode, ensures fast heating, accurate and stable temperature control and good temperature reproducibility. Temperature auto-correction function provided.
7. Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact.
a) High accuracy and reproducibility of the graphite furnace auto-sampler, realizing fully automation of graphite furnace analysis.
b) Fully automated control system
8. Auto-changeover and position optimization of integrated flame/graphite furnace atomizer.
9. Automatic 6-lamp turret. High performance HCL can be used directly, which features high intensity spectral line and low background.
10. Auto-ignition, automatic spectral bandwidth selection.
11. Fully automated wavelength scanning and peaking.
12. Adopting mass flowmeter with program control ensures accurate gas intake volume.
13. Automatic setting of Gas flow
14. Perfect safety protection measures
15. Alarm and automatic safety protection to fuel gas leakage, air pressure deficiency and abnormal flaMain Specification Wavelength range 190-900nm Wavelength accuracy Better than ±0.25nm Resolution Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. Baseline stability 0.004A/30min Background correction The D2 lamp background correction capability at 1A is better than 30 times.
The S-H background correction capability at 1.8A is better than 30 times.Light Source System Lamp turret 6-lamp turret ( Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.); auto alignment; auto fine adjustment; fully automated scan and peak-picking Lamp current adjustment Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. Lamp power supply mode Powered by two pulses (wide and narrow) simultaneously. Optical System Monochomator Single beam, Czerny-Turner design grating monochromator Grating 1800 l/mm Focal length 277mm Blazed Wavelength 250nm Spectral Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.2nm, auto-switch over Atomizer Flame/Graphite Furnace Flame & Graphite Furnace atomizer automatic changeover Automatic graphite furnace position optimization, automatic flame height setting and horizontal position optimization. Automatic gas flow setting, alarm and automatic protection function for gas leak, abnormal flow, air-pressure deficiency and abnormal flame extinction, automatic ignition. Graphite Furnace Autosampler Sample tray capacity 55 sample cups and 5 cups for matrix modifier Material of the sample cup Polypropylene Volume of the standard cup 3ml for sample, 20ml for matrix modifier Number of repeatable measurements Up to 99 times Sampling system 100ul and 5ml injectors, large flow sampling nozzle rinsing. Sampling precision and repeatability Precision: 1%; Repeatability: 0.3% for min. sampling volume. Auto-correction Sampling nozzle auto-correction, sample cup height auto correction. Detection and Data Processing System Detector Photomultiplier with high sensitivity and wide spectral range. Software Windows 98/Me operating system Analytical method Working Curve Method (1st order and 2nd order fitting) with automatic sensitivity correction; Standard Addition Method. Concentration Auto-calculation of concentration and content Repeatable measurement Repeated measurement of up to 99 times, automatic calculation of mean value, standard deviation and relative standard deviation. Multi-task Functions Sequential measurement for multi-element determination to the same sample (Max. 6 elements). Condition reading With model function Result printing Measurement data and final analytical report printout, editing with Excel. Flame Analysis Burner 10cm single slot all-titanium burner Spray chamber All-plastic spray chamber. Nebulizer High efficiency glass nebulizer with metal sleeve. Characteristic Concentration and Detection Limit Air-C2H2 flame
Oxygen-rich Air-C2H2 flameCu: Characteristic concentration≤0.025mg/L,
Detection limit≤0.006mg/L;
Ba: Characteristic concentration ≤ 0.22mg/L
Al: Characteristic concentration ≤ 0.4mg/LGraphite Furnace Analysis Temperature range Room temperature~3000ºC Light-controlled temperature range 1500ºC -3000ºC, temperature auto-correction Temperature control accuracy ≤1% Heating rate 2000ºC/s Characteristic mass Cd≤0.5 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g Precision Cd≤3%, Cu≤3%, Mo≤4% Safety protection Protections for over-heating, over-current, and less cooling water and argon pressure monitoring functions Dimensions and Weight Main unit 107X49X58cm, 140kg Graphite furnace 42X42X46cm, 65kg Autosampler 40X29X29cm, 15kg
- Inquiry About Atomic Absorption Spectrophotometer (WFX-210)* is required.
- Contact Info
-
Company: Tsingtao Unicom-Optics Instruments Co.,Ltd Area: China/Shandong Contact: Mr. Domty Tel: 86-532-88480588 Fax: 86-532-88480587 Web site: www.unicomoptics.com Add: Building 2, Huanghai West Road, Laixi City, Qingdao, Shandong, China Zip: 266600
- Supplier's information
-
Tsingtao Unicom-Optics Instruments Co.,Ltd
Tsingtao Unicom-Optics Instruments Co., Ltd. Tsingtao Unicom-Optics Instrumen ..... more >>
- Other products from this supplier
- View the similar products
-