  • Product Detials:

    Atomic Absorption Spectrophotometer (WFX-210)

    1. Innovated flame analysis technique
    2. The patented flame analysis technique adopting oxygen-rich air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950ºC, which makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost. Oxygen-rich flame will not pollute the environment and is not harmful to human bodies. It is a break-through in flame AAS analysis.
    3. Popular Czerny-Turner grating monochromator with 1800l/mm provides higher resolution.
    4. All plastic pre-mix spray chambers with high anti-corrosion ability.
    5. Accurate and reliable graphite furnace analysis
    6. Light-controlled graphite furnace power supplier, adopting FUZZY-PID temperature control technique and dual curve working mode, ensures fast heating, accurate and stable temperature control and good temperature reproducibility. Temperature auto-correction function provided.
    7. Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact.
    a) High accuracy and reproducibility of the graphite furnace auto-sampler, realizing fully automation of graphite furnace analysis.
    b) Fully automated control system
    8. Auto-changeover and position optimization of integrated flame/graphite furnace atomizer.
    9. Automatic 6-lamp turret. High performance HCL can be used directly, which features high intensity spectral line and low background.
    10. Auto-ignition, automatic spectral bandwidth selection.
    11. Fully automated wavelength scanning and peaking.
    12. Adopting mass flowmeter with program control ensures accurate gas intake volume.
    13. Automatic setting of Gas flow
    14. Perfect safety protection measures
    15. Alarm and automatic safety protection to fuel gas leakage, air pressure deficiency and abnormal fla

    Main SpecificationWavelength range190-900nm
    Wavelength accuracyBetter than ±0.25nm
    ResolutionTwo spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%.
    Baseline stability0.004A/30min
    Background correctionThe D2 lamp background correction capability at 1A is better than 30 times.  
    The S-H background correction capability at 1.8A is better than 30 times.
    Light Source SystemLamp turret6-lamp turret ( Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.); auto alignment; auto fine adjustment; fully automated scan and peak-picking
    Lamp current adjustmentAutomatic adjustment and display.           Wide pulse current: 0~25mA,              Narrow pulse current: 0~10mA.                 
    Lamp power supply modePowered by two pulses (wide and narrow) simultaneously.  
    Optical SystemMonochomatorSingle beam, Czerny-Turner design grating monochromator
    Grating1800 l/mm
    Focal length277mm
    Blazed Wavelength250nm
    Spectral Bandwidth0.1nm, 0.2nm, 0.4nm, 1.2nm, auto-switch over
    AtomizerFlame/Graphite FurnaceFlame & Graphite Furnace atomizer automatic changeover
    Automatic graphite furnace position optimization, automatic flame height setting and horizontal position optimization.
    Automatic gas flow setting, alarm and automatic protection function for gas leak, abnormal flow, air-pressure deficiency and abnormal flame extinction, automatic ignition.
    Graphite Furnace AutosamplerSample tray capacity55 sample cups and 5 cups for matrix modifier
    Material of the sample cupPolypropylene
    Volume of the standard cup3ml for sample, 20ml for matrix modifier
    Number of repeatable measurementsUp to 99 times
    Sampling system100ul and 5ml injectors, large flow sampling nozzle rinsing.
    Sampling precision and repeatabilityPrecision: 1%; Repeatability: 0.3% for min. sampling volume.
    Auto-correctionSampling nozzle auto-correction, sample cup height auto correction.
    Detection and Data Processing SystemDetectorPhotomultiplier with high sensitivity and wide spectral range.
    SoftwareWindows 98/Me operating system
    Analytical methodWorking Curve Method (1st order and 2nd order fitting) with automatic sensitivity correction; Standard Addition Method.
    ConcentrationAuto-calculation of concentration and content
    Repeatable measurementRepeated measurement of up to 99 times, automatic calculation of mean value, standard deviation and relative standard deviation.
    Multi-task FunctionsSequential measurement for multi-element determination to the same sample (Max. 6 elements).
    Condition readingWith model function
    Result printingMeasurement data and final analytical report printout, editing with Excel.
    Flame AnalysisBurner10cm single slot all-titanium burner   
    Spray chamberAll-plastic spray chamber.
    NebulizerHigh efficiency glass nebulizer with metal sleeve.
    Characteristic Concentration and Detection LimitAir-C2H2 flame
    Oxygen-rich Air-C2H2 flame
    Cu: Characteristic concentration≤0.025mg/L,
    Detection limit≤0.006mg/L;
    Ba: Characteristic concentration ≤ 0.22mg/L
    Al:  Characteristic concentration ≤ 0.4mg/L
    Graphite Furnace AnalysisTemperature rangeRoom temperature~3000ºC
    Light-controlled temperature range1500ºC -3000ºC, temperature auto-correction
    Temperature control accuracy≤1%
    Heating rate2000ºC/s
    Characteristic massCd≤0.5 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g
    PrecisionCd≤3%, Cu≤3%, Mo≤4%
    Safety protectionProtections for over-heating, over-current, and less cooling water and argon pressure monitoring functions
    Dimensions and WeightMain unit107X49X58cm, 140kg
    Graphite furnace42X42X46cm, 65kg
    Autosampler40X29X29cm, 15kg

Inquiry About Atomic Absorption Spectrophotometer (WFX-210)* is required.
 
Subject: *
Message: *
 
Company Name: *
Your Name: *
Business Phone: *
(Type: +86-571-88228426)
Fax:
Business Email: *
Zip:
Address:
CheckCode:      *

         
Contact Info
Company: Tsingtao Unicom-Optics Instruments Co.,Ltd
Area: China/Shandong
Contact: Mr. Domty
Tel: 86-532-88480588
Fax: 86-532-88480587
Web site: www.unicomoptics.com
Add: Building 2, Huanghai West Road, Laixi City, Qingdao, Shandong, China