SH100 Humidifier

erkundigen

Kategorie: Medical devices/General Surgery Devices
Erklären: 1. The environment requirement:
Temperature 10 degrees--30 degrees
The relative humility ≤ 80%
Environment pressure 86KPa_106KPa
2. Work pressure
≤ 20KPa.
3. The electrical requirement
The power: AC220V± 22V, 50Hz
Fuse: 0, 5A
Power<85VA
4. The airway temperature test
Sensor type: Temperature sensor
The test range: 20degrees--50degrees
The error: ± 1degrees
5. Weight and size
W: 135mm H: 145mm D: 95mm without chamber
1.2kg without chamber
2.0kg with chamber and water
6. The safety function
It is adaptive for IEC60601-1 1995
When the temperature inside chamber is over 95degrees, the machine stops to heat.
When the temperature of airway is over 38degrees, the machine will stop to heat and at the same time it will produce sound and display alarm signal. When the temperature of airway is low than 32degrees, the machine will produce sound and display alarm signal too.
Firmenname: Beijing Jinxin Technology Co.,Ltd
Region: China/Beijing
Kontakt: Mr. Keven(International Busine
Tel: 86-10-83653457
Fax:
Webseite: http://www.jinxintechnology.com
Adresse: Rm. 29, Building 5, No. 249 East Tie Jiang Ying, Fengtai District, Beijing, China
Postleitzahl: 100075