Medical Ventilator (PA 900B)
Classificazione: |
Medical devices/Other Devices |
Descrizione del prodotto: |
PA-900B Ventilator
Specifications
Mode IPPV, SIPPV, IMV, PCV Respiratory Rate 1-99/min
SIMV Rate 1-20/min Inspiratory Flat 0-1S
I: E 1/4.0-4.0/1 Tidal Volume 50ml-1500ml
Maximum Safe Pressure ≤ 6kPa TV Adjust Coefficient 80-120%
Airway Pressure Upper Limit 1-6kPa
Airway Pressure Lower Limit 0-5kPa
Sigh Volume ≥ 150% of tidal volume Ptr. -1.0-1.0kPa
PEEP 0.1-1.0kPa Alarm Silence ≤ 120S
Display 10.4'' colorful TFT screen Optional Battery ≥ 30min
Oxygen Concentration 21%-100%
Monitor Parameter
Tidal Volume 0-9990mL Airway Pressure 0-10kPa
Air Flow 0-99.9L Total Respiratory Rate 0-99/min
Respiratory Rate 0-99/min Power Supply AC/DC
Airway Resistance; Lung Conformability; Pressure-time Wave Form; Flow-time Wave Form
Other
Air Compressor; IRV
Temperature range of humidifier 37-67
Trademark:
Perlong