Light source
Quartz-halogen lamp 12v/20w
Wavelength range
Automatic by 8 positions filter-wheel
7 standard filters: 340, 405, 492, 510, 546, 578, 630nm
1 position free for optional filter
Half bandwidth < 12nm
Stray light: < 1.0 % ( 340 nm )
Photometric range
0.0000 - 3.0000 abs
Resolution: 0.0001 abs
Drift: 0.002 a/hour
Flowcell
Stainless steel body with quartz window
Measuring volume 32 ul
Aspiration volume programmable: 200 -2000 ul
Optical path 10mm
Carryover <2.0% (500ul)
Analysis method
End point, with or without reagent blank
End point, with sample blank and with or without reagent blank
Kinetic, with or without linearity check
Fixed time, with or without reagent blank
Absorbance
Turbidimetry
Linear and non-linear standardization
Parameter settings
Method
Wavelength
Temperature
Reagent blank y/n
Sample blank y/n
Delay time
Measuring time
Reaction type
Absorbance limit
Aspiration volume
Standards
Linearity check
Unit for results
Memory: 3000 results
Temperature control
By means of peltier elements
25, 30, 37 degree optional
Precision: +/- 0.1
Input: Touch screen
Display: Large lcd display
Printer: Built-in thermal printer
Interface: Rs-232 serial port
Opreating environment
Temperature: 15 -30 degree
Humidity: 20 - 80 %
Atmospheric pressure: 86-106kpa
Power requirement
Ac 230v +/- 10%, 50/60hz
Ac 110v +/- 10%, 50/60hz
Dimension: 420mm x 380mm x 170mm
Weight: 8.5 kg
More Product Features
Trademark:
CE
Model NO.:
NB201
Standard:
CE,
Productivity:
2000pcs/year
Unit Price/Payment:
USD 1770
Company:
Caretium Medical Instruments Co.,Ltd. |
Firmenname: |
Caretium Medical Instruments Co.,Ltd |
Region: |
China/Guangdong |
Kontakt: |
Mr. Bruce Li |
Tel: |
86-755-25273714 |
Fax: |
86-755-25273096 |
Webseite: |
http://www.caretium.com/index_en.asp |
Adresse: |
Beishan Industrial Park 7th Floor Building 1, Beishan Road, Yantian, Shenzhen, Guangdong, China |
Postleitzahl: |
518083 |