Sterilization pouch-flat pouch(heat sealing)
Material: ARJO-WIGGINS
1. To choose pouches as per the size the equipment's which to be sterilized. And then put the clean, dry and equipment into the pouches.
P. S.: The equipment should be less than 3/4 of the sizes of the pouches to ensure the seal successfully and avoid the break of the pouches.
2. Make the acuate edge of the equipment to be contrary to the peel direction when we put the acuate equipment in to the pouches.
3. Seal the pouches with the heat-seal machine, marking the detailed information such as the sterilization date.
4. Put the film side of the pouches against the heat bar of the heat-seal machine, the valid temperature should be 200-230 OC, the valid pressure should be 0.15kg/cm2, and the valid time should be 4--6 seconds.
5. Put the closed pouches into the steam clave or ETO sterilizer, according to the rule to let the pouches' paper side against paper side, and then do the sterilization as per the normal methods.
6. To identify the situation of indicator's color change: If sterilization proceeded, it will turn black from initial blue based on steam sterilization method, and will turn brown from initial pink based on the ETO sterilization method.
7. Store the sterilized pouches dry, clean and dark-light place.
8. Tear the pouches as per the peeling direction and put the sterilized equipment out.