Feasible infusion speed range: 1-1000ml/h (Infusion increment is 1ml/h)
Feasible preset infusion volume limit: 1-9999ml or unlimited
Accumulated infusion volume display: 0-9999ml
Nfusion increment: 1ml/h
Infusion precision: <8%
Intravenous infusion speed: 0.8-1.2ml/h KVO rate
Sensitive of bubble: ≥ 5mm
Occlusion pressure value: 40-150KPa
Alarm function: Occlusion, infusion pipe air bubble, door open, liquid exhaustion, infusion completion, low battery, installation error
Power: 200-240VAC, 49-51Hz
Inside battery: Rechargeable Ni-hydrogen battery 2000mAh, charge 16hours, and discharge time ≥ 90min when the infusion speed is 25ml/h;
Fuse: 2A, 250V (inside, change must be by professional)
Operational condition: Temperature 41-104,
Relative temperature≤ 80%
Storage condition: Environment temperature 14-104,
Pressure 70~106kPa, relative humidity ≤ 80%
Work system: Consecutive work
Maximum size of outer shell: 140(W)*160 (D) *230 (H) mm
Weight: About 2.8 Kg
Applicable infusion pipe: All ordinary infusion pipe of national standards
Screen display: Infusion speed; Preset infusion quantity; Accumulated infusion quantity; Battery capacit; Battery charging indication; AC power indication; Infusion completion; Occlusion alarm; Infusion pipe air bubble etc.
Electromagnetic wave (Hertzian wave) protection: Accord with the stipulation of IEC60601-1-2 (1993 edition)
Electrical safety: Accord with the stipulation of GB 9706.27-2005.
More Product Features
Trademark:
kangertai
Model NO.:
KET-S1
Standard:
1-1000ml/h
Productivity:
50000/year
Unit Price/Payment:
Comes The Electricity Inquiry
Min. Order:
1
Export Markets:
North America, South America, Eastern Europe, Southeast Asia, Africa, Oceania, Mid East, Eastern Asia, Western Europe
Company:
Ningbo Kangertai Medical Instrument Co., Ltd. |