Erklären: |
1. Work condition
1.1 Ambient temperature: -5 ---- 40º C
1.2 Relative humidity: 20% ~ 90%
1.3 Power supply: AC220V(± 22V, 50Hz(± 1Hz
1.4 Internal batteries: 12V, 3000mAh Battery under-voltage≤10.5V)
Battery full charged(24hours)
1.5 Atmospheric Pressure: 860hpa~1060hpa
1.6 Storage Temperature: -40º C~+55º C
2. Basic Parameters
2.1 Dimensions (mm): 270(L)X65(W)X71(H)
2.2 Weight: Approximately 3.2 kg
2.3 Power Waste: ≤20VA
2.4 Compatible Syringe Size: 10ml, 20ml, 30ml, 50ml Disposable syringe
2.5 Infusion Flow Selection:
0.1~199.9ml/h (10ml), 0.1~299.9ml/h (20ml),
0.1~399.9ml/h (30ml), 0.1~599.9ml/h (50ml)
Original value is 0.1ml/h
2.6 Infusion total amount setting: 0.1ml -50 ml
2.7 Block Pressure: 40~70K pa
2.8 Infusion Flow Precision: (± 5%(0.1ml/h~50ml/h)
(± 3%(50ml/h~599.9ml/h)
2.9 Alarm: Self-Test, Block, Near-End, Completion, AC/DC changing, Cutting AC wire, Low-battery, Idle/Move, Toppling alarm (Presetting)
2.10 Functions: Syringe? Type Test, Transfusion Total Amount Display, Transfused Amount Display, Power Direction, KVO Function, Remaining Amount Adjusting, Evacuate Bubble Function, Bolus, Safety Double Confirmation Function, Ethernet RJ-45 Port, Nurse Call Jack.
Model NO.:
LINZ-6-II-B
Origin:
China
Company:
Shanghai Leien Medical Equipment Co., Ltd. |